696 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
32 :WMEMory<r> Commands
:WMEMory<r>:DISPlay
(see page 790)
Command Syntax
:WMEMory<r>:DISPlay <on_off>
<r> ::= 1-2 in NR1 format
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :WMEMory<r>:DISPlay command turns the display of the specified
reference waveform on or off.
There are two reference waveform locations, but only one reference
waveform can be displayed at a time. That means, if :WMEMory1:DISPlay
is ON, sending the :WMEMory2:DISPlay ON command will automatically set
:WMEMory1:DISPlay OFF.
Query Syntax
:WMEMory<r>:DISPlay?
The :WMEMory<r>:DISPlay? query returns the current display setting for
the reference waveform.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also Chapter 32, “:WMEMory<r> Commands,” starting on page 693
":WMEMory<r>:CLEar" on page 695
":WMEMory<r>:LABel" on page 697