Agilent Technologies 2000 X-Series Manual de Serviço Página 448

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 930
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 447
448 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
SPI (Serial Peripheral Interface) triggering— consists of connecting
the oscilloscope to a clock, data (MOSI or MISO), and framing signal.
You can then trigger on a data pattern during a specific framing period.
The serial data string can be specified to be from 4 to 64 bits long.
UART/RS- 232 triggering (with Option 232) — lets you trigger on
RS- 232 serial data.
Reporting the Setup
Use :SBUS<n>? to query setup information for the :SBUS<n> subsystem.
Return Format
The following is a sample response from the :SBUS1? query. In this case,
the query was issued following a *RST command.
:SBUS1:DISP 0;MODE IIC;:SBUS1:IIC:ASIZ BIT7;:SBUS1:IIC:TRIG:TYPE
STAR;QUAL EQU;:SBUS1:IIC:SOUR:CLOC CHAN1;DATA
CHAN2;:SBUS1:IIC:TRIG:PATT:ADDR -1;DATA -1;DATA2 -1
NOTE
Two I2S buses or two SPI buses cannot be decoded on both SBUS1 and SBUS2 at the same
time.