:CHANnel<n> Commands 10
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 223
:CHANnel<n>:IMPedance
(see page 790)
Command Syntax
:CHANnel<n>:IMPedance <impedance>
<impedance> ::= ONEMeg
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:IMPedance command selects the input impedance setting
for the specified analog channel. The only legal value for this command is
ONEMeg (1 M
Ω).
Query Syntax
:CHANnel<n>:IMPedance?
The :CHANnel<n>:IMPedance? query returns the current input impedance
setting for the specified channel.
Return Format
<impedance value><NL>
<impedance value> ::= ONEM
See Also "Introduction to :CHANnel<n> Commands" on page 218