:DIGital<d> Commands 12
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 247
:DIGital<d>:POSition
(see page 790)
Command Syntax
:DIGital<d>:POSition <position>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<position> ::= integer in NR1 format.
The :DIGital<d>:POSition command sets the position of the specified
channel. Note that bottom positions might not be valid depending on
whether digital buses, serial decode waveforms, or the zoomed time base
are displayed.
Query Syntax
:DIGital<d>:POSition?
The :DIGital<d>:POSition? query returns the position of the specified
channel.
If the returned value is "- 1", this indicates there is no space to display the
digital waveform (for example, when all serial lanes, digital buses, and the
zoomed time base are displayed).
Return Format
<position><NL>
<position> ::= integer in NR1 format.
See Also "Introduction to :DIGital<d> Commands" on page 243
Channel Size Position Top Bottom
Large 0-7 7 0
Medium 0-15 15 0
Small 0-31 31 0
NOTE
This command is only valid for the MSO models.