320 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:X1Position
(see page 790)
Command Syntax
:MARKer:X1Position <position> [suffix]
<position> ::= X1 cursor position in NR3 format
<suffix> ::= {s | ms | us | ns | ps | Hz | kHz | MHz}
The :MARKer:X1Position command:
Sets :MARKer:MODE to MANual if it is not currently set to WAVeform
(see ":MARKer:MODE" on page 319).
Sets the X1 cursor position to the specified value.
X cursor units are set by the :MARKer:XUNits command.
Query Syntax
:MARKer:X1Position?
The :MARKer:X1Position? query returns the current X1 cursor position.
This is functionally equivalent to the obsolete :MEASure:TSTArt
command/query.
Return Format
<position><NL>
<position> ::= X1 cursor position in NR3 format
See Also "Introduction to :MARKer Commands" on page 318
":MARKer:MODE" on page 319
":MARKer:X2Position" on page 322
":MARKer:X1Y1source" on page 321
":MARKer:X2Y2source" on page 323
":MARKer:XUNits" on page 325
":MEASure:TSTArt" on page 729
NOTE
If the front-panel cursors are off, the marker position values are not defined and an error is
generated. Make sure to set :MARKer:MODE to MANual or WAVeform to put the cursors in
the front-panel Normal mode.