392 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
21 :MTESt Commands
:MTESt:DATA
(see page 790)
Command Syntax
:MTESt:DATA <mask>
<mask> ::= binary block data in IEEE 488.2 # format.
The :MTESt:DATA command loads a mask from binary block data. These
are the data bytes found in a *.msk file.
Query Syntax
:MTESt:DATA?
The :MTESt:DATA? query returns a mask in binary block data format. The
format for the data transmission is the # definite- length format defined in
the IEEE 488.2 specification.
Return Format
<mask><NL>
<mask> ::= binary block data in IEEE 488.2 # format
See Also ":SAVE:MASK[:STARt]" on page 435
":RECall:MASK[:STARt]" on page 420