Agilent Technologies 2000 X-Series Manual de Serviço Página 155

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 930
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 154
Root (:) Commands 6
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 155
:AUToscale:AMODE
(see page 790)
Command Syntax
:AUToscale:AMODE <value>
<value> ::= {NORMal | CURRent}
The :AUTOscale:AMODE command specifies the acquisition mode that is
set by subsequent :AUToscales.
When NORMal is selected, an :AUToscale command sets the NORMal
acquisition type and the RTIMe (real- time) acquisition mode.
When CURRent is selected, the current acquisition type and mode are
kept on subsequent :AUToscales.
Use the :ACQuire:TYPE and :ACQuire:MODE commands to set the
acquisition type and mode.
Query Syntax
:AUToscale:AMODE?
The :AUToscale:AMODE? query returns the autoscale acquire mode setting.
Return Format
<value><NL>
<value> ::= {NORM | CURR}
See Also "Introduction to Root (:) Commands" on page 150
":AUToscale" on page 153
":AUToscale:CHANnels" on page 156
":ACQuire:TYPE" on page 195
":ACQuire:MODE" on page 187
Vista de página 154
1 2 ... 150 151 152 153 154 155 156 157 158 159 160 ... 929 930

Comentários a estes Manuais

Sem comentários