Obsolete and Discontinued Commands 33
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 735
:MEASure:VSTArt
(see page 790)
Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X1Y1source command.
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also "Introduction to :MARKer Commands" on page 318
"Introduction to :MEASure Commands" on page 340
":MARKer:Y1Position" on page 327
":MARKer:Y2Position" on page 328
":MARKer:YDELta" on page 329
":MARKer:X1Y1source" on page 321
":MEASure:SOURce" on page 361
":MEASure:TDELta" on page 727
":MEASure:TSTArt" on page 729
NOTE
The short form of this command, VSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 792). The normal short form, VST, would be the same for
both VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
NOTE
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y1Position command (see page 327) instead.