514 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
25 :SBUS<n> Commands
:SBUS<n>:UART:BITorder
(see page 790)
Command Syntax
:SBUS<n>:UART:BITorder <bitorder>
<bitorder> ::= {LSBFirst | MSBFirst}
The :SBUS<n>:UART:BITorder command specifies the order of transmission
used by the physical Tx and Rx input signals for the serial decoder and/or
trigger when in UART mode. LSBFirst sets the least significant bit of each
message "byte" as transmitted first. MSBFirst sets the most significant bit
as transmitted first.
Query Syntax
:SBUS<n>:UART:BITorder?
The :SBUS<n>:UART:BITorder? query returns the current UART bit order
setting.
Return Format
<bitorder><NL>
<bitorder> ::= {LSBF | MSBF}
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":SBUS<n>:UART:TRIGger:TYPE" on page 529