Agilent Technologies 2000 X-Series Manual de Serviço Página 388

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 930
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 387
388 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
21 :MTESt Commands
:MTESt:COUNt:FWAVeforms
(see page 790)
Query Syntax
:MTESt:COUNt:FWAVeforms? [CHANnel<n>]
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:COUNt:FWAVeforms? query returns the total number of failed
waveforms in the current mask test run. This count is for all regions and
all waveforms collected on the channel specified by the optional parameter
or collected on the currently specified source channel (:MTESt:SOURce) if
there is no parameter.
Return Format
<failed><NL>
<failed> ::= number of failed waveforms in NR1 format.
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:COUNt:WAVeforms" on page 391
":MTESt:COUNt:TIME" on page 390
":MTESt:COUNt:RESet" on page 389
":MTESt:SOURce" on page 409
Example Code "Example Code" on page 379
Vista de página 387