Obsolete and Discontinued Commands 33
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 729
:MEASure:TSTArt
(see page 790)
Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1
cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also "Introduction to :MARKer Commands" on page 318
"Introduction to :MEASure Commands" on page 340
":MARKer:X1Position" on page 320
":MARKer:X2Position" on page 322
":MARKer:XDELta" on page 324
":MEASure:TDELta" on page 727
":MEASure:TSTOp" on page 730
NOTE
The short form of this command, TSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 792). The normal short form "TST" would be the same for
both TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
NOTE
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X1Position command (see page 320) instead.