:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 459
:SBUS<n>:CAN:SAMPlepoint
(see page 790)
Command Syntax
:SBUS<n>:CAN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :SBUS<n>:CAN:SAMPlepoint command sets the point during the bit
time where the bit level is sampled to determine whether the bit is
dominant or recessive. The sample point represents the percentage of time
between the beginning of the bit time to the end of the bit time.
Query Syntax
:SBUS<n>:CAN:SAMPlepoint?
The :SBUS<n>:CAN:SAMPlepoint? query returns the current CAN sample
point setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also "Introduction to :TRIGger Commands" on page 589
":SBUS<n>:MODE" on page 451
":SBUS<n>:CAN:TRIGger" on page 463