246 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
12 :DIGital<d> Commands
:DIGital<d>:LABel
(see page 790)
Command Syntax
:DIGital<d>:LABel <string>
<d> ::= 0 to (# digital channels - 1) in NR1 format
<string> ::= any series of 10 or less characters as quoted ASCII string.
The :DIGital<d>:LABel command sets the channel label to the string that
follows. Setting a label for a channel also adds the name to the label list
in non- volatile memory (replacing the oldest label in the list).
Query Syntax
:DIGital<d>:LABel?
The :DIGital<d>:LABel? query returns the name of the specified channel.
Return Format
<label string><NL>
<label string> ::= any series of 10 or less characters as a quoted
ASCII string.
See Also "Introduction to :DIGital<d> Commands" on page 243
":CHANnel<n>:LABel" on page 225
":DISPlay:LABList" on page 260
":BUS<n>:LABel" on page 204
NOTE
This command is only valid for the MSO models.
NOTE
Label strings are 10 characters or less, and may contain any commonly used ASCII
characters. Labels with more than 10 characters are truncated to 10 characters.