:CALibrate Commands 9
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 213
:CALibrate:STARt
(see page 790)
Command Syntax
:CALibrate:STARt
The CALibrate:STARt command starts the user calibration procedure.
See Also "Introduction to :CALibrate Commands" on page 207
":CALibrate:PROTected" on page 212
NOTE
Before starting the user calibration procedure, you must set the rear panel CALIBRATION
switch to UNPROTECTED, and you must connect BNC cables from the TRIG OUT connector
to the analog channel inputs. See the User's Guide for details.