Agilent Technologies 2000 X-Series Manual de Serviço Página 688

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 930
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 687
688 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
31 :WGEN Commands
:WGEN:VOLTage
(see page 790)
Command Syntax
:WGEN:VOLTage <amplitude>
<amplitude> ::= amplitude in volts in NR3 format
For all waveforms except DC, the :WGEN:VOLTage command specifies the
waveform's amplitude. Use the :WGEN:VOLTage:OFFSet command to specify
the offset voltage or DC level.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example,
an amplitude of 5 V and an offset of 1 V is the same as a high- level
voltage of 4 V and a low-level voltage of - 1 V.
Query Syntax
:WGEN:VOLTage?
The :WGEN:VOLTage? query returns the currently specified waveform
amplitude.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also "Introduction to :WGEN Commands" on page 665
":WGEN:FUNCtion" on page 667
":WGEN:VOLTage:OFFSet" on page 691
":WGEN:VOLTage:HIGH" on page 689
":WGEN:VOLTage:LOW" on page 690
Vista de página 687
1 2 ... 683 684 685 686 687 688 689 690 691 692 693 ... 929 930

Comentários a estes Manuais

Sem comentários