:TIMebase Commands 28
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 581
:TIMebase:RANGe
(see page 790)
Command Syntax
:TIMebase:RANGe <range_value>
<range_value> ::= time for 10 div in seconds in NR3 format
The :TIMebase:RANGe command sets the full- scale horizontal time in
seconds for the main window. The range is 10 times the current
time- per-division setting.
Query Syntax
:TIMebase:RANGe?
The :TIMebase:RANGe query returns the current full-scale range value for
the main window.
Return Format
<range_value><NL>
<range_value> ::= time for 10 div in seconds in NR3 format
See Also "Introduction to :TIMebase Commands" on page 578
":TIMebase:MODE" on page 579
":TIMebase:SCALe" on page 583
":TIMebase:WINDow:RANGe" on page 586
Example Code
' TIME_RANGE - Sets the full scale horizontal time in seconds. The
' range value is 10 times the time per division.
myScope.WriteString ":TIM:RANG 2e-3" ' Set the time range to 0.002
seconds.
See complete example programs at: Chapter 38, “Programming Examples,”
starting on page 799