604 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
29 :TRIGger Commands
:TRIGger[:EDGE]:REJect
(see page 790)
Command Syntax
:TRIGger[:EDGE]:REJect <reject>
<reject> ::= {OFF | LFReject | HFReject}
The :TRIGger[:EDGE]:REJect command turns the low- frequency or
high- frequency reject filter on or off. You can turn on one of these filters
at a time.
The high frequency reject filter adds a 50 kHz low-pass filter in the
trigger path to remove high frequency components from the trigger
waveform. Use the high frequency reject filter to remove high- frequency
noise, such as AM or FM broadcast stations, from the trigger path.
The low frequency reject filter adds a 50 kHz high- pass filter in series
with the trigger waveform to remove any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
that can interfere with proper triggering.
Query Syntax
:TRIGger[:EDGE]:REJect?
The :TRIGger[:EDGE]:REJect? query returns the current status of the reject
filter.
Return Format
<reject><NL>
<reject> ::= {OFF | LFR | HFR}
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:HFReject" on page 593
":TRIGger[:EDGE]:COUPling" on page 602
NOTE
The :TRIGger[:EDGE]:REJect and the :TRIGger[:EDGE]:COUPling selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:COUPling can change the COUPling setting.