Vista de página 550
:SEARch Commands 26
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 551
:SEARch:SERial:LIN:MODE
(see page 790)
Command Syntax
:SEARch:SERial:LIN:MODE <value>
<value> ::= {ID | DATA | ERRor}
The :SEARch:SERial:LIN:MODE command selects the type of LIN
information to find in the Lister display:
ID — searches for a frame ID.
DATA — searches for a frame ID and data.
ERRor — searches for errors.
Frame IDs are specified using the :SEARch:SERial:LIN:ID command.
Data values are specified using the:SEARch:SERial:LIN:PATTern:DATA
command.
Query Syntax
:SEARch:SERial:LIN:MODE?
The :SEARch:SERial:LIN:MODE? query returns the currently selected mode.
Return Format
<value><NL>
<value> ::= {ID | DATA | ERR}
See Also Chapter 26, “:SEARch Commands,” starting on page 531
":SEARch:SERial:LIN:ID" on page 550
":SEARch:SERial:LIN:PATTern:DATA" on page 552