396 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
21 :MTESt Commands
:MTESt:RMODe
(see page 790)
Command Syntax
:MTESt:RMODe <rmode>
<rmode> ::= {FORever | SIGMa | TIME | WAVeforms}
The :MTESt:RMODe command specifies the termination conditions for the
mask test:
FORever — the mask test runs until it is turned off.
SIGMa — the mask test runs until the Sigma level is reached. This level
is set by the ":MTESt:RMODe:SIGMa" on page 401 command.
TIME — the mask test runs for a fixed amount of time. The amount of
time is set by the ":MTESt:RMODe:TIME" on page 402 command.
WAVeforms — the mask test runs until a fixed number of waveforms are
acquired. The number of waveforms is set by the
":MTESt:RMODe:WAVeforms" on page 403 command.
Query Syntax
:MTESt:RMODe?
The :MTESt:RMODe? query returns the currently set termination condition.
Return Format
<rmode><NL>
<rmode> ::= {FOR | SIGM | TIME | WAV}
See Also "Introduction to :MTESt Commands" on page 379
":MTESt:RMODe:SIGMa" on page 401
":MTESt:RMODe:TIME" on page 402
":MTESt:RMODe:WAVeforms" on page 403
Example Code "Example Code" on page 379