Agilent Technologies 1100 Series Manual de Serviço Página 83

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 318
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 82
1100 Series DAD and MWD Reference Manual 83
Troubleshooting and Test Functions 3
Heater Power At Limit
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has
been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable Causes
Ambient conditions have changed too much during the run, so that
optimum results may not be guaranteed.
Suggested Actions
Verify that the reproducibility of your results is not affected
Expose the detector to more stable ambient conditions.
Vista de página 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88 ... 317 318

Comentários a estes Manuais

Sem comentários