1100 Series DAD and MWD Reference Manual 83
Troubleshooting and Test Functions 3
Heater Power At Limit
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has
been hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable Causes
• Ambient conditions have changed too much during the run, so that
optimum results may not be guaranteed.
Suggested Actions
✔ Verify that the reproducibility of your results is not affected
✔ Expose the detector to more stable ambient conditions.