
310 1100 Series DAD and MWD Reference Manual
A Safety Information
Declaration of Conformity for HOX2 Filter
We herewith inform you that the holmium oxide glass filter (type Hoya HY-1)
(Part No. 79880-22711)
meets the following specification of absorbance maxima positions:
361.0 nm - 418.9 nm - 453.7 nm - 536.7 nm (spectral bandwidth: 2 nm).
Agilent Technologies guarantees the traceability of the specified absorbance
maxima to a National Institute of Standards & Technology (NIST) holmium
oxide solution standard with a lot-to-lot tolerance of ±0.3nm.
The wavelength calibration filter built into the Agilent Technologies UV-visible
detectors
is made of this material and meets these specifications. It is, therefore,
suitable for wavelength calibration of these detectors within the specified
wavelength accuracy of the respective detector over its wavelength range.
___March 20, 2000__
Date
Engineering Manager Quality Manager
Ta b l e 5 4
Wavelength Accuracy
Product No. 79880A HP 1090 system ± 1 nm
79883A HP 1090 system ± 1 nm
79854A HP 1050 series ± 1 nm
1306A HP 1050 series ± 1 nm
G1315A/B Agilent 1100 series ± 1 nm
G1365A/B Agilent 1100 series ± 1 nm
G1600A ± 1 nm
Comentários a estes Manuais