1100 Series DAD and MWD Reference Manual 137
Repairing the Detector 4
6
The assembly may be disassembled using a pair of pliers
to remove the retainer ring. The window may be cleaned or
the complete assembly must be exchanged.
7 Insert the rebuilt or new assembly and slide it in until it is
stopped by the filter lever.
8 Leave slight clearance for movement of the filter lever
between both, the lens assembly and the filter lever, and
fix the setscrew.
Next steps:
Perform a filter test to check the proper function.
Replace the flow cell cover, see “Cleaning or
Exchanging the Holmium Oxide Filter” on page 127.
Replace both lamps.
Replace the front cover.