164 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Exchanging the Power Supply
1 Turn off the lamp.
2 Switch off the detector, and disconnect the cables.
3 Remove the detector from the stack and place it on the working bench.
4 Remove the front cover, top cover and top foam section, see “Removing the
Top Cover and Foam” on page 140.
5 Remove the detector main board, see “Exchanging the Detector Main
Board” on page 143.
6 Remove the fan assembly, see “Exchanging the Fan (G1315A/G1365A)” on
page 148 or “Replacing Fan Parts (G1315B/G1365B)” on page 150.
7 Remove the optical unit, see “Removing the Optical Unit
(G1315A/G1365A)” on page 154 or “Removing the Optical Unit
(G1315B/G1365B)” on page 156.
When required If defective
Tools required Screwdriver POZI 1 PT3
Wrench 1/4 inch
Wrench 5 mm
Wrench 7 mm
Parts required Power supply 0950-2528
NOTE
The repair level of the power supply assembly is exchanging the complete assembly. No
serviceable parts are inside.