1100 Series DAD and MWD Reference Manual 167
Repairing the Detector 4
Exchanging the Leak Sensor
1 Turn off the lamp.
2 Switch off the detector and disconnect the cables.
3 Remove detector from stack and place it on the working bench.
4 Remove the front cover, top cover and top foam section, see “Removing the
Top Cover and Foam” on page 140.
5 Remove the optical unit, see “Removing the Optical Unit
(G1315A/G1365A)” on page 154 or “Removing the Optical Unit
(G1315B/G1365B)” on page 156.
6 Disconnect the leak sensor cable from the detector main board (J21).
When required If defective
Tools required Screwdriver POZI 1 PT3
Parts required Leak sensor assembly 5061-3356
WARNING
Do not switch on the detector when the cover is removed.
CAUTION
Electronic boards and components are sensitive to electronic discharge (ESD). In order
to prevent damage always use an ESD protection (for example, the ESD wrist strap
from the accessory kit) when handling electronic boards and components, see “Using
the ESD Strap” on page 98.
NOTE
On the G1315B DAD and G1365B MWD the leak sensor can be pulled out without the
removing the leak pan (the next steps describe the procedure for the G1315A/G1365A).