50 1100 Series DAD and MWD Reference Manual
3 Troubleshooting and Test Functions
Overview of the Detector’s Indicators and Test Functions
Status Indicators
The detector is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the detector. The status
indicators provide a quick visual check of the operation of the detector (see
page 51).
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the detector
generates an error message in the user interface. For each message, a short
description of the failure, a list of probable causes of the problem, and a list of
suggested actions to fix the problem are provided (see “Error Messages” on
page 53).
Wavelength Recalibration
Wavelength recalibration is recommended after repair of internal
components, and on a regular basis to ensure correct operation of the
detector. The detector uses the deuterium alpha and beta emission lines for
wavelength calibration (see “Wavelength Recalibration” on page 84).
Test Funct io ns
A series of test functions are available for troubleshooting and operational
verification after exchanging internal components (see “Test Functions” on
page 85).