Agilent Technologies 1100 Series Manual de Serviço Página 173

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 318
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 172
1100 Series DAD and MWD Reference Manual 173
Repairing the Detector 4
Installing the Optical Unit (G1315B/G1365B)
When required When repairs have been completed
Tools required Screwdriver POZI 1 PT3
NOTE
Use this procedure for the the G1315B/G1365B only. For G1315A/G1365A detectors refer
to “Installing the Optical Unit (G1315A/G1365A)” on page 171.
CAUTION
The fan must be installed in the correct orientation to ensure correct cooling and
operation of the detector, see “Replacing Fan Parts (G1315B/G1365B)” on page 150.
Preparations for this procedure:
The power supply, the fan and the processor board are
already installed.
1 Check that the shock absorbers and the leak sensor cable
is placed correctly.
Vista de página 172
1 2 ... 168 169 170 171 172 173 174 175 176 177 178 ... 317 318

Comentários a estes Manuais

Sem comentários