Agilent Technologies 1100 Series Manual de Serviço Página 105

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 318
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 104
1100 Series DAD and MWD Reference Manual 105
Repairing the Detector 4
Exchanging a Flow Cell
When required If an application needs a different type of flow cell or the flow cell needs
repair.
Tools required Two 1/4 inch wrenches for capillary connections
Parts required Standard flow cell, 10 mm, 13 µl, 120 bar, G1315-60012
Semi-micro flow cell, 6 mm, 5 µl, 120 bar, G1315-60011
High pressure flow cell, 6 mm, 1.7 µl, 400 bar, G1315-60015
Nano flow cell, refer to “Nano Flow Cell - Replacing or Cleaning” on page 122
Preparations for this procedure:
Turn the lamp(s) off.
1 Press the release buttons and remove the front cover to
have access to the flow cell area.
Vista de página 104
1 2 ... 100 101 102 103 104 105 106 107 108 109 110 ... 317 318

Comentários a estes Manuais

Sem comentários