Agilent Technologies 1100 Series Manual de Serviço Página 60

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 318
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 59
60 1100 Series DAD and MWD Reference Manual
3 Troubleshooting and Test Functions
Leak Sensor Open
The leak sensor in the detector has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes
Leak sensor not connected to the detector main board.
Defective leak sensor.
Suggested Actions
Ensure the leak sensor is connected correctly.
Exchange the leak sensor.
Vista de página 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 317 318

Comentários a estes Manuais

Sem comentários