1100 Series DAD and MWD Reference Manual 285
Control Module Screens for Agilent 1100 DAD and MWD 7
Info Log
Use the m-key to receive a pop-up menu, Select Info Log. A list of the last
events are listed. For troubleshooting reasons they can be printed or saved to
a file on the PCMCIA card.
EMF (Early Maintenance Feedback)
Use the F1 key (EMF) to set EMF parameters. Choose menu item 1 (Setup
limits) to select lamp-on level for the UV and Vis lamp at which you want to
receive a warning.
If a set limit has been exceeded, a message box will pop up. This will not stop a
sequence or run (information only to plan maintenance activities). If you press
Reset, the limits will be removed. Ignore will continue to keep the EMF flag
set.