158 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Exchanging the Source Lens (Achromat) Assembly
When required The source lens (achromat) assembly might be cleaned or replaced when the
transmission of light is reduced and a lamp exchange or new spectro
windows do not improve the light throughput
Tools required Screwdriver POZI 1 PT3
hexagonal key 1.5 mm
A pointed pair of pliers
Parts required Source lens (achromat) assembly G1315-65201(includes filter lever
G1315-45001)
Spectro window 79880-28111
Filter lever G1315-45001
NOTE
If the filter motor was removed, the filter lever should not be reused. Always use a new
filter lever to assure correct fit on the filter motor shaft.
Preparations for this procedure:
Turn off the lamp.
Switch off detector and disconnect the power cable.
Remove detector from stack and place it on the working
bench.
Remove the front cover, top cover and top foam section,
see “Removing the Top Cover and Foam” on page 140.
Remove the optical unit, see “Removing the Optical Unit
(G1315A/G1365A)” on page 154 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
1 Place the optical unit on your bench.