1100 Series DAD and MWD Reference Manual 151
Repairing the Detector 4
The heater and temperature sensor assemblies are
clipped into the fan assembly. In case the
temperature sensor or the heater assembly fails,
they can be unclipped separately from the fan
assembly.
If the fan is defective, unclip both, the temperature
sensor assembly and the heater assembly from the
fan as described in the next steps.
2 To unclip the heater from the fan, use a pair of pliers,
squeeze the clips and push them through the hole. Repeat
this with the three remaining clips.
3 If required, remove the sensor out of its holder. 4 To unclip the sensor holder from the fan, pull it carefully
from its position pushing into the direction the arrow
points to.