Agilent Technologies 1100 Series Manual de Serviço Página 170

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 318
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 169
170 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Replacing Status Light Pipe
When required If part was broken or removed
Tools required Screwdriver POZI 1 PT3
Parts required Status light pipe 5041-8384
Preparations for this procedure:
Remove the front cover and top cover, see “Removing
the Top Cover and Foam” on page 140.
1 The status light pipe is clipped into the top cover.
Next steps:
Replace the top cover, see “Replacing the Foam and the Top Cover” on page 175.
Replace the detector into the stack and reconnect the cables and capillaries.
Turn on the detector.
Vista de página 169
1 2 ... 165 166 167 168 169 170 171 172 173 174 175 ... 317 318

Comentários a estes Manuais

Sem comentários