Agilent Technologies 1100 Series Manual de Serviço Página 115

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 318
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 114
1100 Series DAD and MWD Reference Manual 115
Repairing the Detector 4
Notes
The fittings at the flow cell body are special types for
low dead volumes and not compatible with other
fittings.
When retightening the fittings, make sure that they
are carefully tightened (handtight plus 1/4 turn with
a wrench). Otherwise damage of the flow cell body
or blockage may result.
2 After replacing the outlet capillary, fix it handtight first.
Then do a 1/4 turn with a 4-mm wrench.
3 To replace the inlet capillary, use a 4-mm wrench for the
fitting.
4 Unscrew the cell body from the heat exchanger and the
heat exchanger from the clamp unit.
Outlet capillary
Inlet capillary
Vista de página 114
1 2 ... 110 111 112 113 114 115 116 117 118 119 120 ... 317 318

Comentários a estes Manuais

Sem comentários