1100 Series DAD and MWD Reference Manual 165
Repairing the Detector 4
5 Carefully remove the bottom foam part by sliding it out
towards the rear.
6 Unscrew the power supply at the rear of the module.
7 Press down the power switch light pipe to remove it from
the coupler.
8 Remove the power supply completely.
Re-use the coupler on the new power supply.
The repair level of the power supply assembly is exchange of the complete assembly. No serviceable parts
inside.