286 1100 Series DAD and MWD Reference Manual
7 Control Module Screens for Agilent 1100 DAD and MWD
Firmware Update
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the DAD or MWD. Use the
F5 key (FW Update) to enter the Update section. If you want to update the
resident firmware (together with specific main firmware revisions), select the
a file from the PCMCIA card (RES_n_nn.DLB) and press execute. If you want
to update the main firmware, press F7 key (Transfer) to turn the module into
the resident mode (LED on module should blink yellow).
Use the Esc key to receive Views on the F5 key. Choose System from the
pull-down menu. Use the F3 key (Records) to select the Generic module. In
this screen the resident firmware revision is shown.