Chapter 3 77
HP-IB Command Guidelines
Guidelines for Operation
The Data Functions are set to their default values whenever
the power is cycled on the Test Set
the front-panel PRESET key is selected
the *RST Common Command is received through HP-IB
Changing Attribute Units The AUNits command can be used to
change the Attribute Units setting for selected measurements. Only the
Attribute Units for power and relative level measurements can be
changed. Table 3-4 lists the measurements which have changeable
Attribute Units.
Before changing the Attribute Units for a selected measurement, the
Test Set verifies that all Data Function values can be properly
converted from the current unit-of-measure to the new unit-of-measure.
The following Data Function settings are checked:
the Reference value
•the High Limit
the Low Limit
the Meter’s high end setting
the Meter’s low end setting
the Meter’s interval
Table 3-4 Measurements with Attribute Units That Can
Be Changed
Function Available Attribute Units
TX Power measurement W or DBM
Adjacent Channel Power
LRATio, URATio DB or PCT
LLEVel, ULEVel W or DBM
SINAD measurement DB or PCT
DISTN measurement DB or PCT
SNR measurement DB or PCT