150 Chapter 6
O:\Manuals\E6380A_Progguid\Book\Advoper.fm
Advanced Operations
Status Reporting
Clearing the Event Register
The EVENT register is cleared whenever it is queried or whenever the
Common Command *CLS is sent to the Test Set.
Reading the Enable Register
Example 6-53 Syntax
STATus:QUEStionable:ENABle?
Example 6-54 Example
OUTPUT 714;"STAT:QUES:ENAB?"
ENTER 714;Register_value
Writing the Enable Register
Example 6-55 Syntax
STATus:QUEStionable:ENABle <integer>
Example 6-56 Example
OUTPUT 714;"STAT:QUES:ENAB 256"
Clearing the Enable Register
The ENABLE register is cleared by writing to it with an integer value
of zero.