218 Chapter 7
O:\Manuals\E6380A_Progguid\Book\Mencard.fm
Memory Cards/Mass Storage
Backing Up Procedure and Library Files
Backing Up Procedure and Library Files
Making a backup copy of procedure and library files helps guard
against file loss due to PC card (or battery) failure.
Using the COPY_PL ROM Program
The COPY_PL procedure on the internal ROM will make backup copies
of Tests Subsystem’s Procedure and Library files onto a second SRAM
PC card, and can also initialize an uninitialized PC card. This program
does not make backup copies of Tests Subsystem’s code files, or copy any
type of file to OTP cards.
The COPY_PL procedure is designed for use with Agilent Technologies
software to make backup copies of Agilent supplied, or user-generated
procedure and library files.
To run COPY_PL:
1. Access the SOFTWARE MENU screen.
2. Select the Select Procedure Location: field and choose ROM.
3. Select the Select Procedure Filename: field and select IB_UTIL.
4. Select the Run Test softkey to start the procedure.
5. Select COPY_PL from the list of utilities.
6. Follow the displayed instructions.