182 Chapter 6
O:\Manuals\E6380A_Progguid\Book\Advoper.fm
Advanced Operations
Instrument Initialization
*RST IEEE 488.2 Common Command
The *RST Reset is accomplished by sending the *RST Common
Command to the Test Set through the HP-IB. Table 6-13 lists the fields
which are restored/initialized when the *RST command is received. The
restored state or initialized value is listed with the field’s name. Fields
which are not listed are maintained at their current value. All fields in
the SOFTWARE MENU screen and the TESTS (External Devices)
screen are maintained. The current state/value of the maintained fields
can be ascertained programmatically.
Table 6-13 Screen Fields Restored/Initialized During *RST Reset
(Continued)
INSTRUMENT
CONFIGURE
Screen Fields
PRINTER
CONFIGURE
Screen Fields
TESTS
(Execution
Conditions)
Screen Fields
TESTS
(Printer Setup)
Screen Fields
I/O
CONFIGURE
RF Offset: Off Print Title field
is cleared
Test output
location: Crt
Test output
location: Crt
Save/Recall:
Internal
(Gen)-(Anl):
0.000000
Results output:
All
Results output:
All
Range Hold:
Auto All
If Unit Under
Test Fails:
Continue
Notch Coupl:
None
Test Procedure
Run Mode:
Continuous
RF Display: Freq
RF Chan Std:
MS AMPS
User Def
Base Freq:
800.000000
Chan Space:
30.0000
(Gen)-(Anl):
45.000000
RF Level Offset:
Off
RF In/Out: 0.0
Duplex Out: 0.0
Antenna In: 0.0