24 Chapter 1
O:\Manuals\E6380A_Progguid\Book\Usehpib.fm
Using HP-IB
Overview of the Test Set
Internal Automatic Control Mode
In Internal Automatic Control mode the Test Set’s operation is
controlled by an application program running on the built-in IBASIC
controller. The built-in controller runs programs written in Agilent
Technologies Instrument BASIC (IBASIC), a subset of the BASIC
programming language used on the HP
9000 Series 200/300 System
Controllers. IBASIC is the only programming language supported on
the built-in IBASIC controller.
Similarities Between the Test Set’s IBASIC Controller and
Other Single-Tasking Controllers
The architecture of the IBASIC controller is similar to that of other
single-tasking instrumentation controllers. Only one program can be
run on the IBASIC controller at any given time. The program is loaded
into RAM from some type of mass storage device. Four types of mass
storage devices are available to the Test Set: SRAM PC cards, ROM
memory cards connected to the HP-IB interface, and internal ROM and
RAM. Three types of interfaces are available for connecting to external
instruments and equipment: HP-IB, RS-232, and 16-bit parallel.
Figure 1-2 on page 29 shows how information is routed inside the Test
Set when it is in Internal Automatic Control mode. In Manual Control
mode certain Test Set resources are dedicated to manual operation.
These resources are switched to the IBASIC controller when an IBASIC
program is running. These include the serial interface ports at select
codes 9, 10, and 11, the HP-IB interface at select code 7, the parallel
printer interface ports at select codes 15 and 16, and the display. In
Manual Control mode, front panel information (instrument settings,
measurement results, data input from the DUT) is routed to the display
through the screen control hardware. In Internal Automatic Control
mode the measurement results and data input from the DUT are routed
to the IBASIC controller through a dedicated HP-IB interface. Also, in
Internal Automatic Control mode, the display is dedicated to the
IBASIC controller for program and graphics display. This means
instrument front panels cannot be displayed on the display when an
IBASIC program is running.