Chapter 8 273
IBASIC Controller
PROGram Subsystem
Example 8-16 Example BASIC program to download an IBASIC program to
Test Set
10 OUTPUT 714;"PROG:DEL:ALL" !Delete current program
20 OUTPUT 714;"PROG:DEF #257" !Create program, send header
30 OUTPUT 714;"10 FOR J = 1 TO 10" !18 characters + CR + LF
40 OUTPUT 714;"20 DISP J" !9 characters + CR + LF
50 OUTPUT 714;"30 BEEP" !7 characters + CR + LF
60 OUTPUT 714;"40 NEXT J" !9 characters + CR + LF
70 OUTPUT 714;"50 END"!6 characters
80 END
:DEFine? The :DEFine? query command is used to upload an IBASIC
program from the Test Set to an external controller.
The IBASIC program uploaded to the external controller is transferred
as IEEE 488.2 Definite Length Arbitrary Block Response Data. The
following information describes some of the characteristics of the IEEE
488.2 Definite Length Arbitrary Block Response Data type. Refer to the
IEEE Standard 488.2-1987 for detailed information on this data type.
The data starts with a header which begins with a “#”, followed by a
single non-zero digit in the range 1-9 which specifies the number of
digits in the following count field, followed by a series of digits in the
range of 0-9 which gives the number of data bytes being sent,
followed by the number of data bytes specified by the count field.
Example 8-17 Example
#16<data byte><data byte><data byte><data byte><data byte>
<data byte>
The transfer is terminated by the transmission, from the Test Set to
the external controller, of the response message terminator (NL &
END message).
<NL> = new line = ASCII linefeed character.
<END> = IEEE 488.1 END message.
Example 8-18 Syntax
PROGram[:SELected]:DEFine?