Chapter 1 45
Using HP-IB
IEEE 488.1 Remote Interface Message Capabilities
IEEE 488.1 Remote Interface Message
Capabilities
The remote interface message capabilities of the Test Set and the
associated IEEE 488.1 messages and control lines are listed in Table
1-3.
Table 1-3 Test Set IEE 488.1 Interface Message Capability
Message
Type
Implemented Response IEEE 488.1
Message
Data Yes All front-panel functions, except those listed
in Table 1-2 on page 43, are programmable.
The Test Set can send status byte, message
and setting information. All measurement
results (except dashed “- - - -” displays) and
error messages are available through the bus.
DAB
END
MTA
MLA
OTA
Remote Yes Remote programming mode is entered when
the Remote Enable (REN) bus control line is
true and the Test Set is addressed to listen.
The R annunciator will appear in the
upper-right corner of the display screen when
the Test Set is in remote mode. All front-panel
keys are disabled (except for the LOCAL key,
POWER switch, Volume control and Squelch
control knobs). When the Test Set enters
remote mode the output signals and internal
settings remain unchanged, except that
triggering is reset to the state it was last set
to in remote mode (Refer to “Triggering
Measurements” on page 105).
REN
MLA
Local Yes The Test Set returns to local mode (full
front-panel control) when either the Go To
Local (GTL) bus command is received, the
front-panel LOCAL key is pressed or the REN
line goes false. When the Test Set returns to
local mode the output signals and internal
settings remain unchanged, except that
triggering is reset to TRIG:MODE:SETT
FULL;RETR REP. The LOCAL key will not
function if the Test Set is in the local lockout
mode.
GTL
MLA
Local
Lockout
Yes Local Lockout disables all front-panel keys
including the LOCAL key. Only the System
Controller or the POWER switch can return
the Test Set to local mode (front-panel
control).
LLO