Agilent Technologies 8481B Especificações

Consulte online ou descarregue Especificações para Multímetros Agilent Technologies 8481B. Agilent Technologies 8481B Specifications Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
A
Document Part Number 5971-2668
Printed in the UK September 2004
Errata
This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-Packard's former test
and measurement, semiconductor products and chemical analysis businesses are now part of Agilent
Technologies. To reduce potential confusion, the only change to product numbers and names has been in the
company name prefix: where a product number/name was HP XXXX the current name/number is now Agilent
XXXX. For example, model number HP8648 is now model number Agilent 8648.
Ce manuel peut contenir des références à <<HP>> ou <<Hewlett-Packard.>> Veuillez noter que les produits de
test et mesure, de semi-conducteur et d'analyse chimique qui avaient fait partie de la société Hewlett-Packard
sont maintenent une partie de la société Agilent Technologies. Pour reduire la confusion potentielle, le seul
changement aux noms de reference a été dans le préfixe de nom de société : là où un nom de référence était HP
XXXX, le nouveau nom de référence est maintenant Agilent XXXX. Par example, le HP 8648 s'appelle
maintenent Agilent 8648.
Diese Gebrauchsanweiseung kann Bezug nehmen auf die Namen HP oder Hewlett-Packard. Bitte beachten Sie,
dass ehemalige Betriebsbereiche von Hewlett-Packard wie HP-Halbleiterprodukte, HP-chemische Analysen
oder HP-Test- und Messwesen nun zu der Firma Agilent Technology gehören. Um Verwirrung zu vermeiden
wurde lediglich bei Produktname und - Nummer der vo laufende Firmenname geändert: Produkte mit dem
Namen/Nummer HP XXXX lauten nun mehr Agilent XXXX. Z.B, das Modell HP 8648 heißt nun Agilent 8648.
Questo manuale potrebbe contenere riferimenti ad HP o Hewlett-Packard. Si noti che le attività precedentemente
gestite da Hewlett-Packard nel campo di Test & Misura, Semiconduttori, ed Analisi Chimica sono ora diventate
parte di Agilent Technologies. Al fine di ridurre il rischio di confusione, l'unica modifica effettuata sui numeri di
prodotto e sui nomi ha riguardato il prefisso con il nome dell'azienda : dove precedentemente compariva "HP
XXXX" compare ora "Agilent XXXX". Ad esempio: il modello HP8648 è ora indicato come Agilent 8648.
Este manual puede hacer referencias a HP o Hewlett Packard. Las organizaciones de Prueba y Medición (Test
and Measurement), Semiconductores (Semiconductor Products) y Análisis Químico (Chemical Analysis) que
pertenecían a Hewlett Packard, ahora forman parte de Agilent Technologies. Para reducir una potencial
confusión, el único cambio en el número de producto y nombre, es el prefijo de la compañía: Si el producto solía
ser HP XXXX, ahora pasa a ser Agilent XXXX. Por ejemplo, el modelo HP8648 es ahora Agilent 8648.
5971-2668 Small.fm Page 1 Wednesday, September 22, 2004 2:20 PM
Vista de página 0
1 2 3 4 5 6 ... 153 154

Resumo do Conteúdo

Página 1

ADocument Part Number 5971-2668Printed in the UK September 2004ErrataThis manual may contain references to HP or Hewlett-Packard. Please note that He

Página 2 - マニュアル・チェンジ

List of Related Publicationsviii Agilent E4418B/E4419B Service GuideList of Related PublicationsThe Agilent E4418B and Agilent E4419B User’s Guides ar

Página 3 - Service Guide

Replaceable PartsFront Panel Connector Repair5-6 Agilent E4418B/E4419B Service GuideFront Panel Connector RepairThe front panel assembly is an exchang

Página 4 - Printed in the UK

Replaceable PartsFirmware UpgradesAgilent E4418B/E4419B Service Guide 5-7Firmware UpgradesThe power meter’s firmware can be upgraded using a PC. The c

Página 5 - Legal Information

Replaceable PartsAssembly and Disassembly Guidelines5-8 Agilent E4418B/E4419B Service GuideAssembly and Disassembly GuidelinesThe guidelines in this s

Página 6 - Exclusive Remedies

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-9Removing the A1 Power Supply Assembly1. Remove the power su

Página 7 - Equipment Operation

Replaceable PartsAssembly and Disassembly Guidelines5-10 Agilent E4418B/E4419B Service GuideRemoving the A2 Processor Assembly1. Remove the A5 daughte

Página 8 - Markings

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-11Removing the A4 Comms Assembly1. Remove the A2 processor,

Página 9 - Installation Instructions

Replaceable PartsAssembly and Disassembly Guidelines5-12 Agilent E4418B/E4419B Service GuideRemoving the A5 Daughter or A6 Measurement Assemblies1. Di

Página 10 - List of Related Publications

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-13Removing the A7 Fan Assembly1. Remove the A1 power supply

Página 11 - Sales and Service Offices

Replaceable PartsAssembly and Disassembly Guidelines5-14 Agilent E4418B/E4419B Service GuideRemoving the Power Meter RAM Battery (BT1)1. Remove the A2

Página 12 - United States:

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-15Removing the A3 Front Panel Assembly1. Disconnect the foll

Página 13

Sales and Service OfficesAgilent E4418B/E4419B Service Guide ixSales and Service OfficesFor more information about Agilent Technologies test and measu

Página 14 - About this Guide

Replaceable PartsAssembly and Disassembly Guidelines5-16 Agilent E4418B/E4419B Service GuideReplacing the Front Panel Power Sensor Cable Assemblies1.

Página 15 - Table of Contents

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-174402serv.book Page 17 Tuesday, October 14, 2003 3:18 PM

Página 16

Replaceable PartsAssembly and Disassembly Guidelines5-18 Agilent E4418B/E4419B Service GuideReplacing the Front Panel Power Reference Cable Assembly1.

Página 17

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-19Replacing the Rear Panel Power Sensor Cable Assemblies (Op

Página 18

Replaceable PartsAssembly and Disassembly Guidelines5-20 Agilent E4418B/E4419B Service Guide4402serv.book Page 20 Tuesday, October 14, 2003 3:18 PM

Página 19 - Specifications

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-21Replacing the Rear Panel Power Reference Cable Assembly (O

Página 20

Replaceable PartsAssembly and Disassembly Guidelines5-22 Agilent E4418B/E4419B Service GuideReplacing the Rear Panel Recorder Output(s) Cable Assembly

Página 21 - Power Meter Specifications

Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-23Replacing The Chassis Assembly1. Remove the Power Meter Co

Página 22 - Accuracy

Replaceable PartsAssembly and Disassembly Guidelines5-24 Agilent E4418B/E4419B Service Guide4402serv.book Page 24 Tuesday, October 14, 2003 3:18 PM

Página 23 - 1 mW Power Reference

6Troubleshooting4402serv.book Page 1 Tuesday, October 14, 2003 3:18 PM

Página 24 - Zero Drift of Sensors

Sales and Service Officesx Agilent E4418B/E4419B Service GuideEurope:Agilent TechnologiesTest & MeasurementEuropean Marketing OrganizationP.O. Box

Página 25 - Measurement Noise

TroubleshootingIntroduction6-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter enables qualified service personnel to diagnose suspected f

Página 26

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-3Figure 6-1: Troubleshooting Guide -Upper Window TTL OutputStartPower Up meterPOSTPa

Página 27 - Settling Time

TroubleshootingIntroduction6-4 Agilent E4418B/E4419B Service GuideASelectPreset/localConfirmMeas/SetupLimitsLimits OnTTL OutputTTL Output On, Fail O/P

Página 28 - 12481632641282565121024

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-5Figure 6-2: Troubleshooting Guide - Lower Window TTL OutputStartPower Up meterPOSTP

Página 29

TroubleshootingIntroduction6-6 Agilent E4418B/E4419B Service GuideASelectPreset/localConfirmLower Window - usingMeas/SetupLimitsLimits OnTTL OutputTTL

Página 30 - Power Sensor Specifications

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-7Figure 6-3: Troubleshooting Guide - TTL InputsStartPower Up meterPOSTPass?Connect S

Página 31

TroubleshootingIntroduction6-8 Agilent E4418B/E4419B Service GuideABDid powermeter channelszero?Apply TTL logic 1 level to RJ45 connector. pin 5 wrtpi

Página 32 - General Characteristics

TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-9BRemove A2 Processor Assy (see chapter 5) to gain accessto the Communications Assemb

Página 33 - Environmental Characteristics

TroubleshootingIntroduction6-10 Agilent E4418B/E4419B Service GuideFigure 6-4: Troubleshooting Guide - RS232/422 InterfaceStartPower upmeterPOSTPass?

Página 34 - Dimensions

Measurement Uncertainty Analysis – Instrument Accuracy TestAgilent E4416A/E4417A Service Guide A-1Appendix AMeasurement Uncertainty Analysis – Instrum

Página 35 - Non-Volatile Memory

Sales and Service OfficesAgilent E4418B/E4419B Service Guide xiReturning Your Power Meter for ServiceUse the information in this section if you need t

Página 36

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-2 Agilent E4416A/E4417A Service GuideMeasurement IntroductionIn t

Página 37 - Performance Tests

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-3Uncertainties1. There is no

Página 38

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-4 Agilent E4416A/E4417A Service GuideThe Measurement Uncertainty

Página 39 - Equipment Required

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-5Standard Uncertainty u(xi):T

Página 40 - Test Procedures

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-6 Agilent E4416A/E4417A Service GuideWorst case error on 0.1V ran

Página 41 - Zero Test

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-7Worst case error on 15.8V ma

Página 42 - Procedure

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-8 Agilent E4416A/E4417A Service GuideBy using the DMM Uncertainti

Página 43

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-9Combined Standard Uncertaint

Página 44 - Test Setup

Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-10 Agilent E4416A/E4417A Service Guide4402serv.book Page 10 Tue

Página 45

Measurement Uncertainty Analysis – Power Reference Level TestAgilent E4416A/E4417A Service Guide B-1Appendix BMeasurement Uncertainty Analysis – Power

Página 46

About this Guidexii Agilent E4418B/E4419B Service GuideAbout this GuideChapter 1: SpecificationsThis chapter lists the power meter’s specifications an

Página 47 - Equipment

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-2 Agilent E4416A/E4417A Service GuideMeasurement IntroductionIn

Página 48

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-3where,Ps is the reflection

Página 49

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-4 Agilent E4416A/E4417A Service Guide:The equation above includ

Página 50

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-5A normal distribution is a

Página 51 - Power Reference Level Test

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-6 Agilent E4416A/E4417A Service GuideConsider the model below:A

Página 52

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-7So u(Ms) from the equation

Página 53

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-8 Agilent E4416A/E4417A Service GuideResistanceThe uncertainty

Página 54

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-9Table B-1: Measurement Re

Página 55

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-10 Agilent E4416A/E4417A Service GuideThe readings in Table B-1

Página 56 - Description

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-11Using the above values gi

Página 57

Table of ContentsPageAgilent E4418B/E4419B Service Guide Contents-1Specifications...

Página 58

Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-12 Agilent E4416A/E4417A Service GuideCombined Standard Uncerta

Página 59 - Specification

IndexIndex-1 Agilent E4418B/E4419B Service GuideAA1 power supply/Battery Charger assembly 4-3A2 processor assembly 4-4A3 front panel assembly 4-7A4 in

Página 60

Index-2 Agilent E4418B/E4419B Service Guidefront panel repair 5-6GGP-IBconnector characteristic 1-14Hhumidity characteristic 1-15IIEC 1010-1 1-viiinst

Página 61

Agilent E4418B/E4419B Service Guide Index-3processor assembly 5-10RAM battery 5-14recorder output cable 5-22replaceable parts 5-2returning for service

Página 62

Index-4 Agilent E4418B/E4419B Service Guide4402serv.book Page 4 Tuesday, October 14, 2003 3:18 PM

Página 63

Contents-2 Agilent E4418B/E4419B Service GuideEquipment ... 2-5Test Setup..

Página 64

Agilent E4418B/E4419B Service Guide Contents-3Power Reference Oscillator Level Adjustment... 3-6Description...

Página 65 - Max Actual Result

Contents-4 Agilent E4418B/E4419B Service GuideTroubleshooting... 1Intro

Página 66

1Specifications4402serv.book Page 1 Tuesday, October 14, 2003 3:18 PM

Página 67

マニュアル・チェンジ 変更 本文中の「HP(YHP)」、または「(横河)ヒューレット・パッカード株式会社」という語句を、「Agilent」、または「アジレント・テクノロジー株式会社」と変更してください。 ヒューレット・パッカード社の電子計測、半導体製品、化学分析ビジネス部門は分離独立

Página 68

SpecificationsIntroduction1-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter details the power meter’s specifications and supplemental ch

Página 69 - Adjustments

SpecificationsPower Meter SpecificationsAgilent E4418B/E4419B Service Guide 1-3Power Meter SpecificationsMeterFrequency Range100 kHz to 110 GHz, power

Página 70

SpecificationsPower Meter Specifications1-4 Agilent E4418B/E4419B Service GuideAccuracyInstrumentationAbsolute: ±0.02 dB (Logarithmic) or ±0.5% (Linea

Página 71 - Post-Repair Adjustments

SpecificationsPower Meter SpecificationsAgilent E4418B/E4419B Service Guide 1-51 mW Power Reference1• Power Output: 1.00 mW (0.0 dBm). • Accuracy: (fo

Página 72

SpecificationsPower Meter Supplemental Characteristics1-6 Agilent E4418B/E4419B Service GuidePower Meter Supplemental CharacteristicsMeasurement Speed

Página 73

SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-7Measurement NoisePower sensor dependent (refer to Table 1

Página 74

SpecificationsPower Meter Supplemental Characteristics1-8 Agilent E4418B/E4419B Service GuideTable 1-3: Power Sensor Specifications±Power Sensor Zero

Página 75

SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-9Settling Time0 to 99% settled readings over the GP-IB.For

Página 76

SpecificationsPower Meter Supplemental Characteristics1-10 Agilent E4418B/E4419B Service GuideFor Agilent E-series power sensorsIn FAST mode, within t

Página 77

SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-11Auto filter, default resolution, 10 dB decreasing power

Página 78 - Measurement Uncertainty

Agilent Technologies Part no. E4418-90059Revision CSeptember 2003Service GuideAgilent TechnologiesE4418B/E4419B Power Meters4402serv.book Page i Tue

Página 79

SpecificationsPower Meter Supplemental Characteristics1-12 Agilent E4418B/E4419B Service GuidePower Sensor SpecificationsDefinitionsZero SetIn any pow

Página 80

SpecificationsBattery Option 001 Operational CharacteristicsAgilent E4418B/E4419B Service Guide 1-13Battery Option 001 Operational CharacteristicsThe

Página 81 - Theory of Operation

SpecificationsGeneral Characteristics1-14 Agilent E4418B/E4419B Service GuideGeneral CharacteristicsRear Panel ConnectorsRecorder Output(s)Analog 0-1

Página 82

SpecificationsEnvironmental CharacteristicsAgilent E4418B/E4419B Service Guide 1-15Environmental CharacteristicsGeneral ConditionsComplies with the re

Página 83

SpecificationsGeneral1-16 Agilent E4418B/E4419B Service GuideGeneralDimensionsThe following dimensions exclude front and rear panel protrusions:212.6

Página 84 - A2 Processor Assembly

SpecificationsGeneralAgilent E4418B/E4419B Service Guide 1-17Remote ProgrammingInterfaceGP-IB interface operates to IEEE 488.2. RS-232 and RS-422 inte

Página 85

SpecificationsGeneral1-18 Agilent E4418B/E4419B Service Guide4402serv.book Page 18 Tuesday, October 14, 2003 3:18 PM

Página 86

2Performance Tests4402serv.book Page 1 Tuesday, October 14, 2003 3:18 PM

Página 87 - A3 Front Panel Assembly

Performance TestsIntroduction2-2 Agilent E4418B/E4419B Service GuideIntroductionThe procedures in this chapter test the power meter’s electrical perfo

Página 88 - A4 Comms Assembly

Performance TestsIntroductionAgilent E4418B/E4419B Service Guide 2-3Equipment RequiredTable 2-1 lists all the equipment required for the adjustments a

Página 89 - A5 Daughter Assembly

ii Agilent E4418B/E4419B Service Guide© Copyright 1998, 2003Agilent TechnologiesAll rights reserved. Reproduction, adaptation, or translation without

Página 90 - A6 Measurement Assembly

Performance TestsIntroduction2-4 Agilent E4418B/E4419B Service GuidePerformance Test RecordResults of the performance tests may be tabulated in Table

Página 91

Performance TestsZero TestAgilent E4418B/E4419B Service Guide 2-5Zero Test SpecificationDescriptionAfter the power meter is initially calibrated and z

Página 92

Performance TestsZero Test2-6 Agilent E4418B/E4419B Service GuideProcedureThe following procedure should be performed for the Zero Test. Note The pr

Página 93

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-7Instrument Accuracy Test SpecificationDescriptionThe power meter accur

Página 94

Performance TestsInstrument Accuracy Test2-8 Agilent E4418B/E4419B Service GuideTest SetupFigure 2-2: Instrument Accuracy Test SetupProcedureThe foll

Página 95

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-9Note When switching the range calibrator to STANDBY, allow enough ti

Página 96

Performance TestsInstrument Accuracy Test2-10 Agilent E4418B/E4419B Service GuideTable 2-3: Instrument Accuracy Results (For Indication Only)Note T

Página 97 - Assembly Part Numbers

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-11Test Procedure 2Note This test procedure makes use of a 11683A Rang

Página 98

Performance TestsInstrument Accuracy Test2-12 Agilent E4418B/E4419B Service GuideProcedureThe following procedure should be performed for the Instrume

Página 99 - Power Supply Assembly Types

Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-13Note When switching the range calibrator to STANDBY, allow enough t

Página 100 - Front Panel Connector Repair

Legal InformationAgilent E4418B/E4419B Service Guide iiiLegal InformationNoticeInformation contained in this document is subject to change without not

Página 101 - Firmware Upgrades

Performance TestsInstrument Accuracy Test2-14 Agilent E4418B/E4419B Service Guide14. Set the DC calibrator output to the values shown in Table 2-4. Fo

Página 102 - Replaceable Parts

Performance TestsPower Reference Level TestAgilent E4418B/E4419B Service Guide 2-15Power Reference Level Test DescriptionThis test verifies the power

Página 103

Performance TestsPower Reference Level Test2-16 Agilent E4418B/E4419B Service GuideIf an equivalent measurement system is employed for post-factory te

Página 104

Performance TestsPower Reference Level TestAgilent E4418B/E4419B Service Guide 2-17Equipment• Test power meter: Agilent 432A.• Thermistor mount: Agile

Página 105

Performance TestsPower Reference Level Test2-18 Agilent E4418B/E4419B Service Guide4. Preset the power meter:Press , then on the power meter.Set th

Página 106

Performance TestsAgilent E-Series Power Sensor Interface TestAgilent E4418B/E4419B Service Guide 2-19Agilent E-Series Power Sensor Interface TestThe A

Página 107 - Removing the A7 Fan Assembly

Performance TestsAgilent E-Series Power Sensor Functional Test2-20 Agilent E4418B/E4419B Service GuideAgilent E-Series Power Sensor Functional TestDes

Página 108

Performance TestsAgilent E-Series Power Sensor Functional TestAgilent E4418B/E4419B Service Guide 2-212. Press then .3. On the power meter press ,

Página 109

Performance TestsAgilent E-Series Power Sensor Functional Test2-22 Agilent E4418B/E4419B Service GuideTable 2-6: Functional Test ResultTestChannel A

Página 110

Performance TestsOutput Standing Wave Ratio (SWR) TestAgilent E4418B/E4419B Service Guide 2-23Output Standing Wave Ratio (SWR) Test SpecificationDescr

Página 111

Legal Informationiv Agilent E4418B/E4419B Service GuideLimitation of Warranty The foregoing warranty shall not apply to defects resulting from imprope

Página 112 - Assembly

Performance TestsOutput Standing Wave Ratio (SWR) Test2-24 Agilent E4418B/E4419B Service Guidethermistor mount that can be operated at two different r

Página 113 - (Options 002 and 003)

Performance TestsOutput Standing Wave Ratio (SWR) TestAgilent E4418B/E4419B Service Guide 2-25Equipment• Test power meter: Agilent 432A.• Thermistor m

Página 114

Performance TestsOutput Standing Wave Ratio (SWR) Test2-26 Agilent E4418B/E4419B Service Guide5. Round off the DVM reading to two decimal places. Reco

Página 115 - (Option 003)

Performance TestsOutput Standing Wave Ratio (SWR) TestAgilent E4418B/E4419B Service Guide 2-2717.Round off the DVM reading to two decimal places. Reco

Página 116

Performance TestsOutput Standing Wave Ratio (SWR) Test2-28 Agilent E4418B/E4419B Service Guide25.Calculate factor M using equation 2-11 and record the

Página 117

Performance TestsOutput Standing Wave Ratio (SWR) TestAgilent E4418B/E4419B Service Guide 2-29Table 2-7: Output SWR ResultMax Actual Result1.0611.082

Página 118

Performance TestsPerformance Test Record2-30 Agilent E4418B/E4419B Service GuidePerformance Test RecordModel Agilent E4418B/E4419B Power MeterTested b

Página 119 - Troubleshooting

Performance TestsPerformance Test RecordAgilent E4418B/E4419B Service Guide 2-31Output Standing Wave Ratio Test1N/A___________ ___________1.061, 21.08

Página 120 - Introduction

Performance TestsPerformance Test Record2-32 Agilent E4418B/E4419B Service Guide4402serv.book Page 32 Tuesday, October 14, 2003 3:18 PM

Página 121

3Adjustments4402serv.book Page 1 Tuesday, October 14, 2003 3:18 PM

Página 122

Equipment OperationAgilent E4418B/E4419B Service Guide vEquipment OperationWarnings and CautionsThis guide uses warnings and cautions to denote hazard

Página 123

AdjustmentsIntroduction3-2 Agilent E4418B/E4419B Service GuideIntroduction This chapter describes adjustments and checks which ensure proper performan

Página 124

AdjustmentsIntroductionAgilent E4418B/E4419B Service Guide 3-3Equipment RequiredThe adjustment procedures include a list of recommended test equipment

Página 125

AdjustmentsPower Reference Oscillator Frequency Adjustment3-4 Agilent E4418B/E4419B Service GuidePower Reference Oscillator Frequency AdjustmentNote

Página 126

AdjustmentsPower Reference Oscillator Frequency AdjustmentAgilent E4418B/E4419B Service Guide 3-53. Turn the power reference on by pressing , on th

Página 127

AdjustmentsPower Reference Oscillator Level Adjustment3-6 Agilent E4418B/E4419B Service GuidePower Reference Oscillator Level AdjustmentDescriptionThi

Página 128

AdjustmentsPower Reference Oscillator Level AdjustmentAgilent E4418B/E4419B Service Guide 3-7Note The power meter may be returned to the nearest Agi

Página 129 - Instrument Accuracy Test

AdjustmentsPower Reference Oscillator Level Adjustment3-8 Agilent E4418B/E4419B Service GuideBy substituting equation 3 into equation 1 and manipulati

Página 130 - Measurement Introduction

AdjustmentsPower Reference Oscillator Level AdjustmentAgilent E4418B/E4419B Service Guide 3-93. Connect the equipment as shown in Figure 3-3. The lead

Página 131 - Define the Measurand

AdjustmentsPower Reference Oscillator Level Adjustment3-10 Agilent E4418B/E4419B Service Guide14.Remove the power meter’s cover and adjust A2R90 until

Página 132

AdjustmentsDisplay Brightness and Contrast AdjustmentAgilent E4418B/E4419B Service Guide 3-11Display Brightness and Contrast AdjustmentIntroductionThe

Página 133

Equipment Operationvi Agilent E4418B/E4419B Service GuideGeneral Safety ConsiderationsWARNING Before this instrument is switched on, make sure it ha

Página 134

AdjustmentsDisplay Brightness and Contrast Adjustment3-12 Agilent E4418B/E4419B Service Guide4402serv.book Page 12 Tuesday, October 14, 2003 3:18 P

Página 135

4Theory of Operation4402serv.book Page 1 Tuesday, October 14, 2003 3:18 PM

Página 136 - Sensitivity Coefficients Ci:

Theory of OperationIntroduction4-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter describes how each of the power meter’s assemblies oper

Página 137 - Expanded Uncertainty kuc(y):

Theory of OperationA1 Power Supply/Battery ChargerAgilent E4418B/E4419B Service Guide 4-3A1 Power Supply/Battery ChargerThe A1 power supply/battery ch

Página 138

Theory of OperationA2 Processor Assembly4-4 Agilent E4418B/E4419B Service GuideA2 Processor AssemblyThe processor assembly contains the microcontrolle

Página 139 - Reference Level Test

Theory of OperationA2 Processor AssemblyAgilent E4418B/E4419B Service Guide 4-5The LCD controller on the A3 front panel assembly is configured as a me

Página 140

Theory of OperationA2 Processor Assembly4-6 Agilent E4418B/E4419B Service GuideThe microcontroller circuits that control all the above functions, and

Página 141

Theory of OperationA3 Front Panel AssemblyAgilent E4418B/E4419B Service Guide 4-7A3 Front Panel AssemblyThe front panel assembly is made up of a liqui

Página 142 - Uncertainty Equation

Theory of OperationA4 Comms Assembly4-8 Agilent E4418B/E4419B Service GuideA4 Comms AssemblyThe comms assembly contains the circuitry required for rem

Página 143 - Mismatch

Theory of OperationA5 Daughter AssemblyAgilent E4418B/E4419B Service Guide 4-9A5 Daughter AssemblyThe A5 daughter assembly is loaded vertically into t

Página 144

Equipment OperationAgilent E4418B/E4419B Service Guide viiIEC 1010-1 ComplianceThis instrument has been designed and tested in accordance with IEC Pub

Página 145 - Calibration factor

Theory of OperationA6 Measurement Assembly4-10 Agilent E4418B/E4419B Service GuideA6 Measurement AssemblyThere is one measurement assembly in the Agil

Página 146 - Repeatability

Theory of OperationA6 Measurement AssemblyAgilent E4418B/E4419B Service Guide 4-11The DSP chip controls the logic which sets the chopper driver voltag

Página 147

Theory of OperationA8 Rechargeable Battery Assembly4-12 Agilent E4418B/E4419B Service GuideA8 Rechargeable Battery Assembly The A8 Rechargeable Batter

Página 148 - Sensitivity Coefficients (Ci)

Theory of OperationA8 Rechargeable Battery AssemblyAgilent E4418B/E4419B Service Guide 4-13Remove this page and replace with pullout.fm.4402serv.book

Página 149

Theory of OperationA8 Rechargeable Battery Assembly4-14 Agilent E4418B/E4419B Service Guide4402serv.book Page 14 Tuesday, October 14, 2003 3:18 PM

Página 150 - Expanded Uncertainty

5Replaceable Parts4402serv.book Page 1 Tuesday, October 14, 2003 3:18 PM

Página 151

Replaceable PartsIntroduction5-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter contains details of some of the higher level components a

Página 152

Replaceable PartsAssembly Part NumbersAgilent E4418B/E4419B Service Guide 5-3Assembly Part NumbersReference DesignatorDescriptionAgilent Part NumberA1

Página 153

Replaceable PartsAssembly Part Numbers5-4 Agilent E4418B/E4419B Service GuideA304 Front Panel Assembly for:Agilent E4419B Option 003New E4419-61002Ref

Página 154

Replaceable PartsAssembly Part NumbersAgilent E4418B/E4419B Service Guide 5-5Power Supply Assembly TypesBefore ordering an A1 Power Supply Assembly, c