60 Chapter 2
O:\Manuals\E6380A_Progguid\Book\Measrslt.fm
Methods For Reading Measurement Results
RMB ‘MAV’ Example Program
Comments for Recommended Routine
Table 2-2 Comments for Measure Function from MAV
Example Program
Program
Line
Number
Comments
50 Send a Selected Device Clear (SDC) to the Test Set to put the
HP-IB subsystem into a known state. This allows the control
program to regain programmatic control of the Test Set if it is
in an error state when the program begins to run.
60 Command the Test Set to abort the currently executing
measurement cycle. This will force the Test Set to stop waiting
for any measurement results to be available from
measurements which may be in an invalid state when the
program begins to run.
290 Turn event initiated branches off (except ON END, ON
ERROR and ON TIMEOUT) to ensure that the Measure
function will not be exited until it is finished.
300 Set up a 5 second time-out for any I/O activity on the HP-IB.
This will allow the function to recover if the bus hangs for any
reason. The length of the time-out will be implementation
dependent.
310 Set the triggering mode to single followed by a trigger
immediate command. This ensures that a new measurement
cycle will be started when the TRIG:IMM command is sent.
This sequence, that is: set to single trigger and then send
trigger command, guarantees that the measurement result
returned to the ENTER statement will accurately reflect the
state of the DUT when the TRIG:IMM command was sent. The
’IMM’ keyword is optional.
320 Send the query command passed to the Measure function to
the Test Set.
330 Establish a start time against which to compare the
measurement result time-out value passed to the Measure
function.
340 Start the status byte polling loop.
350 Allow the Test Set some time (100 milliseconds) to process the
measurement. When polling the Test Set the polling loop must
give the Test Set time to process the requested measurement.
Since HP-IB command processing has a higher system priority
within the Test Set than measurement functions, constantly
sending HP-IB commands will result in longer measurement
times.