106 Chapter 5
O:\Manuals\E6380A_Progguid\Book\Trigmeas.fm
Triggering Measurements
Triggering Measurements
Trigger Modes
The Trigger Mode is defined by two parameters: retriggering and
settling.
Retriggering
Retriggering refers to what a measurement does once it has completed
a measurement cycle. There are two options:
1. Single retriggering causes the measurement cycle to stop once a
valid measurement result has been obtained. A valid trigger
command must be received to start the measurement again. When a
measurement cycle is completed, the values for all active
measurements are held until another trigger command is received.
This allows the control program to query a group of measurements
that were triggered at the same time. This is the same functionality
as the front-panel HOLD function.
When the trigger mode is set to single retriggering, consecutive
queries of the same measurement (with no intervening trigger
event) will return the same value. Measurements that rely on
external signals or hardware-generated events (such as the DTMF
Decoder) must be re-armed with a new trigger command before
another measurement can be made.
2. Repetitive retriggering causes the measurement cycle to
immediately start over once a valid measurement result has been
obtained. No trigger event must be received to start the
measurement again. Repetitive retriggering will cause
measurements that rely on external signals or hardware generated
events (such as the DTMF Decoder) to be
re-armed upon completion of a measurement cycle (a valid
measurement result has been obtained). When the trigger mode is
set to repetitive retriggering, consecutive queries of the same
measurement return new measured values.
NOTE
If a measurement cycle does not successfully obtain a valid
measurement result, it will continue to try until it does or the
measurement trigger is aborted. This is true for both retriggering
modes.