Agilent Technologies 8935 Series Manual do Utilizador Página 159

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 329
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 158
Chapter 6 159
Advanced Operations
Status Reporting
Hardware Status Register #1 Group
The Hardware Status Register #1 Group contains information about
the Test Set’s hardware. This status group is accessed using the
STATus commands. The Hardware Status Register #1 Group uses
16-bit registers and includes a Condition Register, Transition Filters,
an Event Register, an Enable Register, and a Summary Message. Refer
to the “Status Reporting Structure Overview” on page 121 section for a
discussion of status register operation. Figure 6-13 shows the structure
and STATus commands for the Hardware Status Register #1 Group.
Figure 6-13 Hardware Status Register #1 Group
0
:STATus: HARDware1: ENABle
?
<integer>
:STATus: HARDware1: EVENt ?
:STATus: HARDware1
:PTRansition
:NTRansition
?
<integer>
?
<integer>
:STATus: HARDware1: CONDition ?
Summary Message Bit
Condition
Register
Transition
Filter Registers
Event
Register
Enable
Register
1
2
3
4
5
6
7
8
9
10
11
12
13
14
L
o
g
i
c
a
l
O
R
15
ch4drw13.drw
(to bit 0 of Status
Byte Register)
Vista de página 158