Agilent Technologies 8935 Series Manual do Utilizador Página 112

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 329
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 111
112 Chapter 5
O:\Manuals\E6380A_Progguid\Book\Trigmeas.fm
Triggering Measurements
Triggering Measurements
Arming Hardware-Triggered Measurements
Some measurements require an external signal to trigger the
measurement. These measurements require that the measurement be
“armed” in order for it to be triggered by the external signal. The
:TRIGger:IMMediate command is used to arm these types of
measurements within the Test Set.
When the trigger mode is set to RETRigger SINGle, the measurement
must be re-armed after each measurement cycle.
When the trigger mode is set to RETRigger REPetitive, the
measurement is continually re-armed after each measurement cycle.
NOTE
Bus Lock Up:
If the required triggering signal is not received, or if the signal level is
incorrect, the measurement will not trigger and the measurement cycle
will not complete. If a measurement cycle does not successfully obtain a
valid measurement result, it will continue to try until it does (an
external trigger is detected) or until the measurement trigger is
aborted. This is true for both retriggering modes. This has the
consequence that both the HP-IB bus and the Active Controller
handshake are in a temporary holdoff state while the Active Controller
waits to read the measurement result from the Test Set.
The control program should include measurement time-out routines
that CLEAR the bus and ABORt the trigger if a measurement does not
complete within a specified amount of time. This provides a method of
preventing the bus from remaining in the temporary holdoff state
indefinitely.
Vista de página 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 328 329

Comentários a estes Manuais

Sem comentários