Agilent Technologies 8935 Series Manual do Utilizador Página 107

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 329
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 106
Chapter 5 107
Triggering Measurements
Triggering Measurements
Settling
Settling refers to the amount of delay introduced to allow signal
transients to propagate through the analysis chain and settle out.
There are two options:
1. Full settling introduces the appropriate delay for all signal
transients which might have occurred at the front panel at just the
same time as the trigger event, to pass through the analysis chain
and settle out. Delays are also inserted to allow for internal
hardware transients to settle.
2. Fast settling introduces no delay for internal or external signal
transients to settle. The programmer must account for transient
settling before issuing the Trigger Event.
NOTE
There will still be delays introduced by autoranging. If you want to
remove these delays, turn OFF autoranging and explicitly set the
ranging amplifiers in the program. Delays introduced by the
measurement processes themselves cannot be eliminated.
NOTE
Bus Lock Up:
If a measurement cycle does not successfully obtain a valid
measurement result, it will continue to try until it does or until the
measurement trigger is aborted. This is true for both retriggering
modes. This has the consequence that both the HP-IB bus and the
Active Controller handshake are in a temporary holdoff state while the
Active Controller waits to read the measurement result from the Test
Set.
The control program should include measurement time-out routines
that CLEAR the bus and ABORt the trigger if a measurement does not
complete within a specified amount of time. This provides a method of
preventing the bus from remaining in the temporary holdoff state
indefinitely.
Vista de página 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 328 329

Comentários a estes Manuais

Sem comentários