Chapter 3 85
Troubleshooting
Manual Troubleshooting Procedures (Step 4)
To isolate the RF Source:
1. On the Test Set:
a. Press
Preset.
b. Press the
Inst Config key to access the INSTRUMENT
CONFIGURE screen.
•Set the RF Display field to Freq.
•Set the RF Offset field to Off.
c. Press the
RF Gen key (to go to the analog RF GENERATOR
screen).
d. Set RF Gen Freq to 1800 MHz.
e. Set Amplitude to 0 dBm.
f. Set Output Port to Dupl.
2. On the external RF modulation analyzer or spectrum analyzer:
a. Set the tuning for the signal generated by the Test Set.
b. Connect the analyzer’s input to the Test Set’s DUPLEX OUT
connector.
3. Set the Test Set’s RF Gen Freq to 1800, then 600, 300, and 150 MHz.
For each frequency, the external RF analyzer should read as follows:
a. Power should read approximately 0.001 W for each frequency.
b. Frequency should read 1800, 600, 300, and 150 MHz respectively.