Chapter 8 221
Performance Tests
CDMA Generator Amplitude Level Accuracy Performance Test 25
CDMA Generator Amplitude Level Accuracy
Performance Test 25
The amplitude level accuracy of the CDMA generator is measured
directly with a power meter. These measurements are made at the top
and bottom of the CDMA generator’s vernier range.
Setup 1
Figure 8-30 Setup 1 for Measurements to 1 GHz
Procedure 1
The following procedure applies to Setup 1 (shown in Figure 8-30 on
page 221).
1. Before connecting the equipment, on the measuring receiver:
a. Reset the instrument.
b. Zero and calibrate the sensor module.
NOTE
Make sure the sensor module’s calibration data is entered into the
measuring receiver.
2. Connect the equipment as shown in Setup 1.
3. On the measuring receiver:
a. Set the measurement mode to RF Power.
b. Set the display to log.
DUPLEX OUT
Measuring Receiver
INPUT 50Ω
SENSOR
Sensor Module